Accession Number : AD0733046


Title :   An Improved Automated Method for Plasma Glucose Analysis with High Specificity and Sensitivity in the 50 to 50 Microgram Range,


Corporate Author : ARMED FORCES RADIOBIOLOGY RESEARCH INST BETHESDA MD


Personal Author(s) : Sobocinski,P. Z. ; Hartley,K. M. ; Canterbury,W. J.


Report Date : OCT 1971


Pagination or Media Count : 9


Abstract : An improved highly sensitive automated procedure for the analysis of plasma glucose in ultramicro sample volumes is presented. The procedure utilizes the glucose oxidase-peroxidase-dye reaction with an extract of gun guaiac as the chromogenic oxygen acceptor. Details for flow manifold construction and reagent composition are presented together with data concerning the precision of the method. (Author)


Descriptors :   (*GLUCOSE, MICROANALYSIS), (*BLOOD CHEMISTRY, GLUCOSE), BLOOD PLASMA, CHEMICAL ANALYSIS, TEST METHODS, AUTOMATION


Subject Categories : BIOCHEMISRTY